Control Module Update
Step 1: Downloading the firmware
Find available versions here R2 firmware files.
Step 2: Updating the Firmware
1. Start the R2. Make sure the R2 is in Ready Mode
.
2. Press F1
to access the Menu.
3. Press the Heating +/- Buttons
to navigate to the USB Mode
tab. Press the Fan/Drum + Button
to enter the USB Mode Settings.
4. On the USB for RoasTime
tab, press the Fan/Drum + Button
. The text will change to USB for SD Card.
5. Connect the R2 to the computer. The drive should show up in Windows Explorer / Finder. If the window does not appear, reconnect the USB Cable. Copy the file to the Firmware folder on the SD card drive and delete the old one.
USB drive connected
You will see a new drive appear on your computer when the R2 is connected in USB Mode. The drive will contain a folder named
Firmware
with a file named bullet_io.bin
. Copy the downloaded firmware file into the firmware folder, replacing the existing file.Verify the filename
Verify the downloaded firmware file is named
If you've downloaded multiple copies, the file might be named
bullet_io.bin
before copying it to the SD card. Do not copy other files into the firmware folder. See the images below.If you've downloaded multiple copies, the file might be named
bullet_io (1).bin
or similar. Please rename the file to bullet_io.bin
before proceeding.Connected R2 drive example on MacOS
R2 firmware folder contents on MacOS.
Replace
bullet_io.bin
with the new firmware file you downloaded.
Connected R2 drive in Windows Explorer
R2 firmware folder contents in Windows Explorer. Replace
bullet_io.bin
with the new firmware file you downloaded.